1. トップ・製品情報
  2. ファイバ製品
  3. Scintillators
  4. Slected Performance Data

Slected Performance Data

Faceplate consists of fused 10-micron fibers in a structure 50mm square by 6mm thick, with 12% statistical EMA

Relative Output of Various Configurations at Different Energies

Plate Type 25kV 50kV 100kV 150kV
Raw Data (ADU)
Scintillating Faceplate 98 164 483 671
Scintillating Faceplate with Mirror on Input 154 258 761 1057
Scintillating Faceplate with Lanex on Input 168 360 851 1115
Hybrid Scint. Faceplate w/ GADOX:Eu Phosphor 556 1143 2175 2468
Lanex Fine Intensifier Screen 260 444 759 811
Normalized Data
Scintillating Faceplate: Basis of Normalization 1.00 1.00 1.00 1.00
Scintillating Faceplate with Mirror on Input 1.57 1.57 1.58 1.58
Scintillating Faceplate with Lanex on Input 1.71 2.20 1.76 1.66
Hybrid Scint. Faceplate w/ GADOX:Eu Phosphor 5.67 6.97 4.50 3.68
Lanex Fine Intensifier Screen 2.65 2.71 1.57 1.21

Comparative Resolution Data (Measured at 50kV)

LP/mm, Max.
Scintillating Faceplate 18
Scintillating Faceplate with Mirror on Input 19
Scintillating Faceplate with Lanex on Input 15
Hybrid Scint. Faceplate w/ GADOX:Eu Phosphor 13
Lanex Fine Intensifier Screen 14

Source of Data: Dr. Clifford Bueno, Lockheed Martin

Absorption Efficiency as a Function of Energy

Fiber Optic GOS Screen
Energy(keV) 12mm thick(%) 15mm thick(%)
50 100 38
100 99 33
150 97 27
200 95 23
250 93 20
300 90 18
350 87 17
400 83 15

Source: Dr.Clifford Bueno, Lockeed, Martin

Comparative Conversion Efficiency

Light Photons per Absorbed X-Ray Photon
Screen Type @50kV @20kV
Lanex 1634 486
Min-R 590 208
Scintillating Fiber Optic 88 33

Source: Dr.Hans Roehrig, University of Arizona

Other Characteristics

  • Core Material: LKH-6 glass, 1.58 refractive index
  • Clad Material: Borosilicate glass, 1.48 refractive index
  • Effective Z of core glass = 31
  • Emission Peaks: 545 & 555 nanometers
  • Reversible radiation browning after 107 rads